並列タイトル等Development and Evaluation of Multi-step Dose Evaluation Method in X-ray Computed Tomography
タイトル(掲載誌)令和3(2021)年度 科学研究費補助金 基盤研究(C) 研究成果報告書 = 2021 Fiscal Year Final Research Report
一般注記A study was conducted with the aim of easily and accurately acquiring the patient dose information required before and after CT imaging. In order to improve the estimation accuracy of size-specific dose estimate (SSDE) before CT scan, it was necessary to apply the correction method to each cross section in the body axis direction. In addition, the dose data of each organ of the patients was analyzed, and a positive correlation was found between the equivalent dose and SSDE in all the organs included in the acquisition range. In order to estimate the organ doses in detail after CT imaging, a dose calculation program corresponding to CT was developed using the Monte Carlo simulation code and it was shown that the absorbed doses can be calculated with high accuracy. Furthermore, improvements have been made to make the dose calculation program compatible with various types of dual-energy CT.
CTの撮影前および撮影後のそれぞれで必要とされる被ばく線量情報を,簡便かつ精度よく評価するための研究を行った.CT撮影前のsize-specific dose estimate(SSDE)の推定精度向上のためには,体軸方向における各横断面で補正手法を適用する必要があった.また,各臓器の線量データを解析し,その等価線量とSSDEとの間には正の相関が認められた.CT撮影後に各臓器の線量を詳細に推定するために,CTに対応した線量計算プログラムを開発し,高い精度で線量が計算できることを示した.さらに,この線量計算プログラムを各種方式のdual-energy CTに対応できるよう改良を進めた.
研究課題/領域番号:18K07746, 研究期間(年度):2018-04-01 – 2022-03-31
出典:研究課題「CT撮影における多段階線量評価法の開発と評価」課題番号18K07746(KAKEN:科学研究費助成事業データベース(国立情報学研究所)) (https://kaken.nii.ac.jp/ja/report/KAKENHI-PROJECT-18K07746/18K07746seika/)を加工して作成
一次資料へのリンクURLhttps://kanazawa-u.repo.nii.ac.jp/?action=repository_action_common_download&item_id=51136&item_no=1&attribute_id=26&file_no=1
関連情報https://kaken.nii.ac.jp/search/?qm=30507372
https://kaken.nii.ac.jp/grant/KAKENHI-PROJECT-18K07746/
https://kaken.nii.ac.jp/ja/report/KAKENHI-PROJECT-18K07746/18K07746seika/
連携機関・データベース国立情報学研究所 : 学術機関リポジトリデータベース(IRDB)(機関リポジトリ)