Jump to main content
規格・テクニカルリポート類

ITER真空容器内中性子モニターのためのマイクロフィッションチェンバーの性能試験

Icons representing 規格・テクニカルリポート類

ITER真空容器内中性子モニターのためのマイクロフィッションチェンバーの性能試験

Material type
規格・テクニカルリポート類
Author
山内 通則ほか
Publisher
-
Publication date
2002-03
Material Format
Digital
Capacity, size, etc.
-
NDC
-
View All

Notes on use

Note (General):

出版タイプ: NAA micro-fission chamber and a dummy chamber without uranium were fabricated and the performance was tested. They are designed to be installed...

Detailed bibliographic record

Summary, etc.:

A micro-fission chamber and a dummy chamber without uranium were fabricated and the performance was tested. They are designed to be installed inside t...

Search by Bookstore

Holdings of Libraries in Japan

This page shows libraries in Japan other than the National Diet Library that hold the material.

Please contact your local library for information on how to use materials or whether it is possible to request materials from the holding libraries.

other

  • JOPSS:JAEA Originated Papers Searching System

    Digital
    You can check the holdings of institutions and databases with which Institutional Repositories DataBase(IRDB)(Institutional Repository) is linked at the site of Institutional Repositories DataBase(IRDB)(Institutional Repository).

Bibliographic Record

You can check the details of this material, its authority (keywords that refer to materials on the same subject, author's name, etc.), etc.

Digital

Material Type
規格・テクニカルリポート類
Author/Editor
山内 通則
西谷 健夫
落合 謙太郎
森本 裕一
堀 順一
海老澤 克之
河西 敏
Publication Date
2002-03
Publication Date (W3CDTF)
2002
Alternative Title
Performance test of micro-fission chambers for in-vessel neutron monitoring of ITER
Periodical title
JAERI-Tech 2002-032
Target Audience
一般
Note (General)
出版タイプ: NA
A micro-fission chamber and a dummy chamber without uranium were fabricated and the performance was tested. They are designed to be installed inside the vacuum vessel of compact ITER (ITER-FEAT) for neutron monitoring. Vacuum leak rate of the chamber, resistances between central conductor and outer sheath, and mechanical strength up to 50G acceleration were confirmed to meet design criteria. Gamma-ray sensitivity was measured with $^{60}$Co gamma-ray irradiation facility at JAERI Takasaki. The output signals for gamma-rays in Campbelling mode were estimated to be less than 0.1\% those by neutrons at the location behind the blanket module in ITER-FEAT. Detector response for 14 MeV neutrons was investigated with the FNS facility. Excellent linearity between count rates and neutron fluxes was confirmed. According to the test for the change of surrounding materials, the sensitivity was enhanced by slow-downed neutrons, which agreed with the calculation result by MCNP-4C code. As a result, it was concluded that the developed micro-fission chamber is applicable for ITER-FEAT.
ITER(ITER-FEAT)真空容器内の中性子モニターの開発を目的として、12mgの二酸化ウランを用いたマイクロフィッションチェンバーとウランのないダミーチェンバーを製作し、性能試験を行った。基本性能として、MIケーブルを取り付けたダミーチェンバーの真空リーク率、チェンバー内の導体と外側容器の絶縁性能、50Gまでの加速度に対する耐性はいずれも設計要求条件を満たした。$\gamma$線に対する感度試験は日本原子力研究所高崎研究所の$^{60}$Coガンマ線照射装置によって行った。それによれば、ITER-FEATブランケット背後の環境で、$\gamma$線に対する感度は中性子に対する感度の0.1\%以下と評価できた。また14MeV中性子に対する検出器の応答は東海研究所の核融合中性子源(FNS)によって試験した。その結果、20$^{\circ}C$(室温)から250$^{\circ}C$までの範囲で計数率と中性子束の良好な直線性が確認できた。遮蔽体がある場合の検出器応答は遮蔽ブランケットの模擬体を用いて試験を行い、MCNP計算の結果と良く一致したデータが得られた。それによると中性子の減速により検出器の感度は上昇するが、遮蔽体の変動による感度の変化は小さい。結論として、本マイクロフィッションチェンバーはITER-FEATの中性子モニターとして充分な性能を有することがわかった。